Volume : 13, Issue : 11, November – 2026
Title:
DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF TOFACITINIB DIHYDROCHLORIDE IN PURE FORM AND PHARMACEUTICAL TABLET DOSAGE FORM
Authors :
Sai Sharanya*, G. Deepika, Vijaya Kuchana, Pasupuleti Sunitha, G. Kalyani
Abstract :
An Analytical, accurate, rugged, rapid and precise reverse phase high performance liquid chromatographic method has been developed for the validated of Tofacitinib Dihydrochloride in bulk form and as well as in pharmaceutical dosage forms. Chromatography was carried out on a Symmetry C18 (4.6mm x 150mm, 5µm) column using a mixture of Methanol and water (45:55% v/v) as the mobile phase at a flow rate of 1.0ml/min, the detection was carried out at 260nm. The retention time of the Tofacitinib Dihydrochloride was 2.368 ±0.02min respectively. The method produce linear responses in the concentration range of 24-120mg/ml of Tofacitinib Dihydrochloride. The method precision for the determination of assay was below 2.0%RSD. The method is useful in the quality control of bulk and pharmaceutical formulations. The developed method was validated for accuracy, precision, linearity, robustness, ruggedness and LOD & LOQ of standard solution. The developed RP-HPLC method was found to be accurate, precise, linear, and robust and was successful applied to a pharmaceutical tablet formulation for qualitative estimation of Tofacitinib Dihydrochloride in Bulk form and Marketed Pharmaceutical Dosage forms.
Keywords: Tofacitinib Dihydrochloride, RP-HPLC, Method Development, Validation, Accuracy.